17 Specifications
198 SRM-3006 Narda
17.4.2 Uncertainty
1 Typical value k = 2 (k = extrapolation or correction factor for calculating the assessment value);
+15 °C to +30 °C
17.4.3 General specifications
Compliance
Dimensions and weight
Extended
measurement
uncertainty
1)
(in conjunction with
SRM-3006 basic
unit and
1.5 m RF cable)
Frequency Range Single axis
measurement with
isotropic antenna
Isotropic
measurement
0.3 – 30 MHz 2.1 dB 2.4 dB
> 30 – 60 MHz 2.2 dB 2.5 dB
> 60 – 250 MHz 2.3 dB 3.2 dB
Calibration
uncertainty
< 1.5 dB
Operating temperature range -10 °C to +50 °C (same as SRM-3006 basic unit)
Air humidity < 29 g/m
3
(< 93 % at +30 °C)
Climatic Storage 1K3 (IEC 60721-3),
extended to -10 °C to +50 °C
Transport 2K4 (IEC 60721-3)
Operating 7K2 (IEC 60721-3)
Mechanical Storage 1M3 (IEC 60721-3)
Transport 2M3 (IEC 60721-3)
Operating 7M3 (IEC 60721-3)
ESD and EMC EN 61326:2006
Safety EN 61010-1:2004
EU Guidelines 2003/11/EG 06.02.2003 (PBDE and OBDE)
2002/95/EG 27.01.2003 (RoHS)
2002/96/EG 27.01.2003 (WEEE)
CE (European Community) Ye s ( Declaration of Conformity on page 205)
Country of origin Germany
Dimensions Length: 450 mm
Antenna head diameter: 120 mm
Weight 470 g