17 Specifications
188 SRM-3006 Narda
17.1.3 Measurement functions
Detection of Narda
measurement antennas
Automatic consideration of antenna parameters after
antenna is plugged in:
• antenna type
• serial number
• calibration date
• antenna factors (see below)
Automatic limitation of the frequency range according to
the frequency range of the connected antenna
Antenna factors Used for display in field strength units
Saved in all Narda antennas during calibration
Antenna factor lists for antennas from other manufacturers
can be saved (these lists defined using the PC
configuration software SRM-3006 Tools)
Detection of Narda
Cables
Automatic consideration of cable parameters after cable is
plugged in:
• Cable type
• serial number
• calibration date
• loss factors (see below)
Automatic limitation of the frequency range according to
the frequency range of the connected cable
Cable loss factors Used for compensation of the power level display
Saved in all Narda cables during calibration
Cable loss lists for cables from other manufacturers can be
saved (these lists defined using the PC configuration
software SRM-3006 Tools included in delivery)
Units • dBV/m
•dBmV/m
•dBA/m
•dBµV/m
•dBV/m
•dBmV/m
•dBμV/m
•dBA/m
•V/m*
•A/m*
•W/m
2*
•mW/cm
2*
• % (of the standard)*
* (with Narda-Antenna only)