12 Spatial Averaging
Narda SRM-3006 147
12.2.2 Discrete
The measurement is controlled manually by pressing the Add Value key.
Each time you press the key, the SRM-3006 performs just one
measurement, confirms this with a beep, and increments the value of
“No. of SAVG” by one. The SRM-3006 averages all the measured values
obtained in this way. The order in which you measure the various points in
the room therefore does not matter if the field is constant over time.
Intermediate and final results can be saved at any time by pressing the
SAVE key.
Figure 31: Discrete recording of three measured values at each of three points in the
room
* Basic standard for the in-situ measurement of electromagnetic field
strength related to human exposure in the vicinity of base stations.
Example application for Discrete Averaging:
EN 50492:2008*