5 The Main Menu
Narda SRM-3006 57
5.2.2 Editing device-specific settings
Main Menu • Settings
To change the settings:
1. Use the rotary control to highlight the item and select it by pressing OK.
2. Use the rotary control to change the setting and press OK to confirm the
change or press ESC to cancel the changes.
3. Press ESC to return to the Main menu.
Table 6: Description of device-specific settings
Item Description and available options
GPS Format GPS data display format:
DegDec (-)ddd.ddddd
MinDec (-)ddd°mm.mmm
DMS deg°mm’ss” N(S,E,W)
Remote Interface Interface used for remote control:
Optical
•USB
Playback Level Earphone playback volume control:
In steps from 1% to 100%
Brightness LCD screen brightness
In steps from 1% to 100%
Power Save (Backlight
switches off after):
Power save mode, backlight switches off automatically:
In steps from 1 min. to 30 min
On permanently
MR Search on Setup
Recall
Search for measurement range when a setup is recalled:
•On
•Off
MR Search Type Measurement range search type:
Normal: The measurement range is selected so that no
saturation occurs in a normal environment.
Conservative: The measurement range is selected so
that no saturation occurs even in an environment with
extremely variable signal strengths (multi-frequency
environment).
Beep on new
Maximum
Beep when new maximum detected:
•Yes
•No