12 Spatial Averaging
150 SRM-3006 Narda
To pause the measurement:
6. Press the Pause softkey.
ª The instrument stops averaging (No. of SAVG is static) and displays
the averaged result. It continues measuring in the background
(indicated by No. of Runs) but does not use these values for
determining the average.
7. If required, move to a new position in the room and press the Continue
softkey.
ª The instrument resumes averaging (indicated by No. of SAVG).
8. Move the antenna along the next path to be measured in the room.
9. You can repeat steps 6 through 8 as often as necessary until you have
the final result.
To save the result in the measurement data memory:
Press the SAVE key.
To delete the result:
Press the Clear softkey.
To exit from Spatial Averaging:
Press the ESC key.
Note: The instrument saves the actual (current) result of spatial averaging
in the measurement data memory when you press the SAVE key.
You can therefore use this function to save intermediate results.
When you press the Continue softkey, the instrument resumes
averaging and takes account of all the measured values obtained so
far. The instrument does not clear all the measurement values
recorded and start a new spatial averaging procedure until you press
the Clear softkey.