12 Spatial Averaging
152 SRM-3006 Narda
Note: The measuring antenna must not be moved during the
measurement, i.e. from the time you press the Add Value key until
the beeper sounds.
5. Move the antenna to the next position in the room to be measured and
press the Add Value softkey.
ª The number of averaged values (No. of SAVG) now indicates 2.
6. Repeat the process for all of the points in the room that are to be
recorded.
To save the result in the measurement data memory:
⇒ Press the SAVE key.
To delete the result:
⇒ Press the Clear softkey.
To exit from Spatial Averaging:
⇒ Press the ESC key.
Note: The instrument saves the actual (current) result of spatial averaging
in the measurement data memory when you press the SAVE key.
You can therefore use this function to save intermediate results.
When you press the Continue softkey, the instrument resumes
averaging and takes account of all the measured values obtained so
far. The instrument does not clear all the measurement values
recorded and start a new spatial averaging procedure until you press
the Clear softkey.