17 Specifications
Narda SRM-3006 185
Safety Evaluation mode
Measurement principle Spectrum analysis, followed by integration over
user-defined frequency bands (services)
Resolution bandwidths RBW
(-3 dB)
Automatic (Auto), or
depending on the narrowest user-defined service
bandwidth, or
user-defined (Manual) for all services, or
separately defined for each individual service
(Individual)
Measurement range setting
(MR)
Set individually from a list or
using the MR Search function for determining the
optimum measurement range at a given time
Detection Root mean square value (RMS),
RMS (integration time 1/(RBW)
Filter See Spectrum Analysis mode
Result Type See Spectrum Analysis mode
Marker functions for bar graph
view
Delta marker on one Result Type or for displaying
the difference between two Result Types
Highest peak, next peak right, next peak left, next
higher peak, next lower peak
Marker field (frequency, level, service name
according to the selected service table)
Evaluation Function Distribution
Axis Isotropic measurement (for direct display of the
isotropic result)
Measurement in the direction of the X, Y, and Z
axis (separate measurement in one direction
using an isotropic / three axis measuring
antenna)
Display Table view showing service names, field
strengths, RBW and the corresponding frequency
band (up to three columns)
Individual screen arrangement
Sort function according to various criteria
Bar graph of services showing contribution of
different Result Types
Noise suppression Identifies whether measured values are above
the device noise floor by setting a threshold
(selectable at 0, 3, 6, 10, 15, or 20 dB relative to
device noise floor).
Measurement values below the threshold are
shown as the absolute threshold value marked
with < (less than threshold)
Others On/Off Measurement of services and gaps in the Service
Table (Others On) or
Measurement of services in the Service Table
excluding gaps (Others Off)