APPLIED OPERATIONS/Protocol processor
IV.
OPERATION INSTRUCTIONS
IV .
2-161
Page
Offset
Offset
ch value
Sets the Offset value of channel.
<E.g.> Enter “Offset 2, 5” to set 5 to the offset
value of channel 2.
PasteROI
PasteROI
Pastes ROI to the next track.
Note: Before the paste, Normal scan is
executed one time.
Pause
Pause
Suspends executing a protocol.
ReceiveCOM
ReceiveCOM
message
Receives RS-232C command from other PC.
The following data returns after command is
executed.
Completes normally: “OK” + Received
characters
Ends with error: “NG” + Received characters.
For further details, see 2-4-7.
Save Save Saves whole series of images.
SendCOM
SendCOM
message [ ,0/1]
Sends data through RS-232C to other PC.
The following data has to be returned in
synchronous mode,.
Completed normally: “OK” + sent characters
Ends with error: “NG” + sent characters
For further details, see 2-4-7.
Size
Size
X, Y
Sets the image size in pixel.
<E.g.> Enter “Size 512, 512” to set 512X512 to
the image size.
Note: The images acquired before and after
executing the Size command cannot be
saved together. In the previous or
succeeding track, execute the Save,
Store, or Stop&Save command to save
the images acquired.
Note) Image size can be set at the following
combination. Be careful that other
combination than the following cannot
be entered. (256, 256) (320, 240) (512,
512) (640, 480) (800, 600) (1024, 768)

Table of Contents