APPLIED OPERATIONS/ Protocol processor
IV.
OPERATION INSTRUCTIONS
IV .
2-162
Page
(1024, 1024) (2048, 2048).
Stop&Save
Stop&Save
Saves all the images acquired before the
Stop&Save command is executed.
Note: The images are saved separately from
the images acquired in succeeding
tracks.
Store Store Opens a new [Display] panel to display the
image.
Zlock
Zlock
[ON/OFF] It sets Valid/Invalid of Z motor lock. Note that
fine tuning handle and coarse tuning handle
cannot be manually operated when Z motor is
locked. In case that manual operation is
required by releasing Z motor lock, difference
of Z stage position displayed on [Zstage] sub
panel of [Acquire] panel of FV300/500 may
occur. Use the handle after confirming the
moving range adequately. In case that Z stage
is controlled with use of command, lock Z
motor. Do not turn fine tuning handle of
microscope in the state that Zlock command is
turned ON as it may cause failure.
<E.g> In case that Z motor lock is required,
enter [Zlock ON].
Zoom
Zoom
value
Sets the zoom ratio.
<E.g.> Enter “Zoom 5” to set 5 to the zoom
ratio.
The images acquired in different sizes cannot be saved together.
Execute the Stop&Save command to save the image acquired before
changing the image size.
One Point!
Multiple image files saved using the Stop&Save command can be combined into a single
time-lapse image file.
Combine the file by referring to section 2-6-6-1, "Appending two images"
NOTE