1. Setup ABL90 FLEX reference manual
1-84
Sample counter
The sample counter lets you keep track of measurements, calibrations and QC.
Press
Utilities > Sample counter to enter the program.
Element Function
Parameter and
Count
List the parameters and how many times each has been
measured by the analyzer. Normally the count is the same
as the total number of measurements if the parameters have
not been excluded from the measurement(s).
Shows the number of sample measurements, calibrations
and QC measurements made since the sample counter was
last reset ("User" column). The following is registered:
Activity Number of...
Total Completed sample/QC measurements/
calibrations only. Interrupted or aborted activities
are excluded.
Aborted Aborted sample/QC measurements/calibrations
due to sample errors, wet-section errors, etc. –
interrupted activities excluded.
Counters
User All completed sample/QC measurements/
calibrations performed by all operators since the
sample counter was last reset.
User counters
last reset
Gives the date when the counters in the "User" column were
last reset to zero.
Buttons Reset counters resets the counters in the "User" column
(on analyzers with no logon protection of the Setup
programs).
Print prints out information in Counters and in Parameter.
Purpose
Description