ABL90 FLEX reference manual 1. Setup
1-29
Step Action
1. Activate the Mandatory temperature check button
2. Or highlight the default temperature in the "Default temperature"
box, enter a default temperature on the keypad and confirm with
Enter.
NOTICES:
 A will appear next to an empty temperature box on the
Quality control identification screen during each QC
measurement; otherwise the result cannot be retrieved.
 The value in C or F is automatically entered on the Quality
control identification screen during measurement. The
temperature can be changed for a particular measurement but
will return to the default setting for future measurements.
QC statistics
In this program you can select the following:
 The statistics factor
 Automatic printing of QC statistics when the lot changes
Step Action
1. Key in the desired statistics factor (from 1.0 to 9.9) on the keypad
and confirm with Enter. The default value is 1.5.
2. Activate the check button in the “Built-in QC” field to automatically
make a printout of the QC statistics if the lot is changed.
3. Press Close to exit.
NOTICE: Statistics factor expands the control range to the statistics range (it is
the range within which QC results must fall in order to be included in the QC
statistics).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals