1. Setup ABL90 FLEX reference manual
1-28
Updating control ranges for all parameters of the displayed level:
Step Action
1. Display the desired slot, using Next slot.
2. Press Update all.
3. Press Continue to update the control ranges for all parameters
under the specified slot, or press Cancel to cancel updating.
NOTICE: Once the Fixed SD has been activated, you cannot update the control
ranges to limits that are narrower than those determined by the Fixed SD, for
both single-parameter and multiple-parameter updates.
QC input setup
In this program you can select the following for the Quality control
identification screen during a manual QC measurement:
Mandatory temperature entry by the operator
The default temperature always displayed (unless changed by the operator)