1. Setup ABL90 FLEX reference manual
1-50
Barometer setup
In this program you can adjust the automatic barometer in accordance with the
reference barometer in your laboratory.
Step Action
1. Key in the desired pressure value on the keypad.
2. Confirm with Enter. The value will be shown in the "Measured
adjusted" box.
Maximum accepted correction is ±19 mmHg (i.e. the difference
between the "Measured unadjusted" and "Measured adjusted"
settings).
Barometer pressure limits are 450-800 mmHg, or 60.0-106.7 kPa,
or 450-800 Torr.
The units are selected in the Setup program: Units.