ABL90 FLEX reference manual 1. Setup
1-23
Quality control setup
Press Menu > Utilities > Setup > QC to access the Quality control solutions
setup and activate a button to enter a program.
The following programs are available:
QC solutions
QC schedule
QC ranges
QC input setup
QC statistics
Westgard Rules
RiLiBÄK ranges
Built-in QC solution is set up together with the Calibration schedule (see earlier
in this chapter)
Manual quality control (QC) solutions
In this program you can assign or change a QC solution to a specific slot –
manual QC measurements only.
Built-in QC results are assigned to the slots A, B and C.
Step Action
1. Highlight a slot, using the arrow buttons.
2. Solutions from Radiometer (the QUALICHECK5+ control solution):
scan the barcode or press Keyboard to enter the barcode
information (see Barcode reader in section Hardware, chapter 2:
What is what in the ABL90 FLEX operator's manual)
Non-Radiometer control solutions: press Add Non-R-.
3. To delete a control solution, highlight the desired slot and press
Delete to cancel the operation.
A warning that this will irreversibly delete all statistical data related
to the selected slot appears. Press Delete to delete the control
solution.
Program