ABL90 FLEX reference manual 1. Setup
1-21
Sample age evaluation setup
This program allows you to set up a maximum sample age for the individual
parameters to enable an automatical sample age evaluation.
Press Menu > Utilities > Setup > Analysis setup > Sample age evaluation
setup.
To enable the sample age evaluation of the individual parameters, do the
following:
Step Action
1. Press the check button in the "Enable sample age evaluation" box.
2. Select the maximum sample age in minutes for pH, using the arrow
buttons.
3. To enable the same number of minutes for all parameters, press the
check button next to "Same rule for all the parameters".
4. Press Close to return to the main screen.
To edit the maximum sample age, do the following:
Step Action
1. In the Parameter/Aging timetable select the desired parameter.
2. In the "Maximum sample age in minutes" box select the desired
number of minutes, using the arrow buttons.
3. Press Close to return to the main screen.