1. Setup ABL90 FLEX reference manual
1-20
Sample pre-registration setup
This program allows you to select interpretation of the barcode and the patient
data that can be confirmed before and during a sample measurement.
Press Menu > Utilities > Setup > Analysis setup > Sample pre-
registration to access the program.
To select the settings, do the following:
Step Action
1. Use the up/down arrow buttons to select the interpretation of the
barcode setting in the "Interpret barcode input as" box.
Choose one of the following:
 Patient ID
 Accession Number
 Sampler ID.
Note that choosing the Accession Number or Sampler ID will gray
out its check button (Sampler ID on the screen above).
2. Select the barcode entry.
3. Activate the relevant check buttons in the" Included fields" box:
Accession no., Patient first name, Patient last name, Birth date,
Patient Sex.
4. Press Close to confirm the settings and return to the main screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals