RADIOMETER ABL90 FLEX - Performance Test Results - Fmethb; Reference Method; Bias Prim.ref

RADIOMETER ABL90 FLEX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7. Performance characteristics ABL90 FLEX reference manual
7-16
ctHb
(g/dL)
FCOHb
(%)
Bias
Sec.ref
S
0
S
X
CV
X
% TE
A
TE
A
(%)
15 0.0 0.00 0.07 0.23 - 0.45 -
15 5.0 0.08 0.07 0.26 5.2 0.59 11.8
15 10.0 0.04 0.06 0.34 3.4 0.71 7.1
15 20.0 0.11 0.07 0.67 3.4 1.42 7.1
15 30.0 0.17 0.07 0.68 2.3 1.50 5.0
15 50.0 0.30 0.08 0.68 1.4 1.63 3.3
15 99.0 0.54 0.12 0.72 0.7 1.95 2.0
Performance test results – FMetHb
Spectrometry, modified Evelyn-Malloy method [Ref. 7].
Setup:
Adult samples. HbF correction is not activated.
ctHb
(g/dL)
FMetHb (%) Bias
Prim.ref
N
15 0 0.12 45
15 20 -0.76 45
N = number of measurements on several analyzers used for the test.
Bias
ABL90-Prim.ref
= Bias
ABL90-ABL735
+ Bias
ABL735-Prim.ref
ctHb
(g/dL)
FMetH
b (%)
Bias
Sec.ref
S
0
S
X
CV
X
% TE
A
TE
A
(%)
15 0.0 -0.04 0.10 0.23 - 0.49 -
15 5.0 0.02 0.09 0.25 5.0 0.51 10.2
15 10.0 -0.04 0.12 0.33 3.3 0.69 6.9
15 20.0 -0.18 0.08 0.27 1.4 0.71 3.5
15 30.0 -0.26 0.08 0.34 1.1 0.93 3.1
15 50.0 -0.21 0.09 0.43 0.9 1.05 2.1
15 99.0 0.11 0.05 0.61 0.6 1.31 1.3
Bias
Sec.ref
and
Repeatability –
blood samples
Reference
method
Bias
Prim.ref
Bias
Sec.ref
and
Repeatability –
blood samples