1. Setup ABL90 FLEX reference manual
1-4
Analyzer security
The Analyzer security program allows you to do the following:
 Set up the general security (set automatic logoff time)
 Add/remove operators from the operator list
 Assign and define access profiles (assign an anonymous operator, i.e. allow
the use of the analyzer without a password, define access to menus for each
access profile, Select configuration of six buttons on the main screen)
This program allows you to hand over the control of the operators and
passwords to the RADIANCE system and to allow an anonymous use of the
analyzer and to define the logoff time of an operator.
To enter this program, press Menu > Utilities > Setup > Analyzer Security
> General Security.
To hand over the control of the operators and passwords to the RADIANCE
system, activate the check button next to "Enable centralized User
Management".
With this option enabled you can only view the operators, not add, edit or
remove any of them. All users defined on the ABL90 FLEX analyzer are deleted
and the list of users in the RADIANCE system is copied to the ABL90 FLEX
analyzer.
NOTICE: This option can only be enabled if RADIANCE communication is
enabled in the RADIANCE Connection Setup program.
To define, how the user should log on, use the up/down arrow buttons in the
"Authenticate" box to select the desired logon option. The following options are
available:
 User ID/Password as primary
This option allows you to enter or scan a User name and password in the
Logon screen. By pressing the Log On BC button a Logon-barcode can be
scanned.
 User ID/Password only
This option allows you to enter or scan the user name and password in the
Logon screen.
 Logon-barcode as primary
This option allows you to enter or scan a Logon-barcode in the Logon screen.
By pressing the Extended Log On button a user name and password can be
scanned.
Programs
General
security

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing