ABL90 FLEX reference manual 8. Parameters
8-7
Is used for both blood and expired air samples.
Definition Partial pressure (or tension) of oxygen in blood.
High and low pO
2
values of arterial blood indicate
blood hyperoxia and hypoxia, respectively.
Unit mmHg; kPa; Torr
Range of indication mmHg; Torr: 0.0-800
kPa: 0.00-107
Reference range mmHg: 83-108 (m, f)
kPa: 11.07-14.40 (m, f)
Conversion of units p (kPa) = 0.133322 × p(mmHg) = 0.133322 ×
p(Torr)
p(mmHg) = p(Torr) = 7.500638 × p(kPa)
Definition Ambient barometric pressure (p(amb)).
Unit mmHg; kPa; Torr
Range of indication mmHg; Torr: 450-800
kPa: 60.0-106.7
Reference range -
Conversion of units p (kPa) = 0.133322 × p(mmHg) = 0.133322 ×
p(Torr)
p(mmHg) = p(Torr) = 7.500638 × p(kPa)
Definition Concentration of total hemoglobin in blood.
Total hemoglobin includes all types of hemoglobin:
deoxy-, oxy-, carboxy-, met-.
Unit g/dL; g/L; mmol/L
Range of indication g/dL: –0.48-27.7
g/L: –4.8-277
mmol/L: –0.30-17.2
Reference range g/dL: 13.5-17.5 (m); 12.0-16.0 (f)
g/L: 135-175 (m); 120-160 (f)
mmol/L: 8.4-10.9 (m); 7.4-9.9 (f)
Conversion of units
ctHb (g/dL) = 1.61140 ctHb (mmol/L);
ctHb (g/L) = 16.1140 ctHb (mmol/L);
ctHb (mmol/L) = 0.62058 ctHb (g/dL) =
0.062058 ctHb (g/L)
Default value 9.3087 mmol/L, (15.0 g/dL or 150 g/L)
pO
2
Baro
ctHb

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals