ABL90 FLEX reference manual 1. Setup
1-73
Mandatory temperature No QC input setup
Default temperature 25 °C
QC schedule Built-in QC (S9030, S9040, S9050)
04:00, 12:00,
20:00
Westgard Rules All rules are "Off"
Run QC after
replacement
On
Program Item Default setting
Inlet gasket 3 months Replacement
schedule
Inlet probe Never
User activities None –
Maintenance
planning
None –
Number of activities before
replacement warning
5
Time before replacement warning 4 hours
Replacement
warnings
Expected measurements per day –
Parameter setup default settings:
Para-
meter
Enabled/
Locked
Repression
Offset
Slope
Units Out-of-range
suppression
pH Not altered No 0.000
1.000 N/A
pCO
2
Not altered No 0.0 1.000 mmHg N/A
pO
2
Not altered No 0.0 1.000 mmHg N/A
ctHb Not altered No N/A 1.000 g/dL No
sO
2
Not altered No 0.0 1.000 % No
FO
2
Hb Not altered No N/A N/A % No
FCOHb Not altered No 0.0 N/A % No
FMetHb Not altered No 0.0 N/A % No
FHbF Not altered No 0 1.000 % Yes
FHHb Not altered No N/A N/A % No
cK
+
Not altered No 0.0 1.000 mmol/L N/A
cNa
+
Not altered No 0 1.000 mmol/L N/A
cCa
2+
Not altered No 0.00 1.000 mmol/L N/A
cCl
Not altered No 0 1.000 mmol/L N/A
cGlu Not altered No 0.0 1.000 mmol/L N/A
cLac Not altered No 0.0 1.000 mmol/L N/A
ctBil Not altered No 0 1.000 µmol/L Yes
Replacement
setup
General setup