ABL90 FLEX reference manual 1. Setup
1-47
Analyzer settings
Press Menu > Utilities > Setup > General setup > Analyzer settings and
activate a button to enter a program.
The following programs are available:
 Analyzer ID
 Time/Date
 Acoustic signal
 Barometer
 Language
Analyzer identification
In this program you can change the analyzer's identification.
Step Action
1. Touch and highlight the "Analyzer name" box if not already
highlighted.
2. Type in an identification name and/or number for the analyzer (up
to 32 characters), using the screen keypad or keyboard. Confirm
with Enter.
NOTICE: The installation number cannot be changed. Quote this number in any
technical inquiries you may have to Radiometer.
Program

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing