5. Sensors and optical system ABL90 FLEX reference manual
5-8
Measurement
A blood sample gives a potential reading of 4.8 mV at the pH sensor. Reading
off from the calibration line shown below, this potential corresponds to a pH of
7.35.
To compensate for deviations from ideal behavior (ex. dilution of sample with
residual rinse solution, and change in gas level by contact with the sample
path), a correction is applied, to give the final value.
The correction is typically a linear correction, and is described for each sensor
type in the following.
Sample
measurements
Corrections

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing