10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-2
List of analyzer messages
The following messages will be seen on the user and manager levels. The
messages are listed in numerical order.
Operator actions are listed in order of priority. Perform the first action in the list;
if unsuccessful, try the next action, etc.
In case of analyzer error or malfunction, the error will be logged in the Activity
log.
Do the following:
Step Action
1. Open the Activity log.
2. Find the relevant error.
3. Highlight it by touching the screen.
4. Press Troubleshoot.
5. Follow the procedures given to remedy the error.
The following table describes possible errors and how to remedy them.
NOTICE: The list contains all possible errors and may thus list errors that will
not be relevant for all analyzer variants. Furthermore, operator actions are in
relation of the analyzer and may differ from local procedures in your institution.
In that case, follow local procedures.
Messages on
user and
manager levels

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals