ABL90 FLEX reference manual 1. Setup
1-65
Corrective actions
In this program you can select the following:
Corrective actions for the events listed in the "Conditions" box
Traffic light signal, if available, for an event
Analyzer action for the subsequent measurements
Step Action
1. Highlight the desired condition, using the up/down arrows in the box.
2. Select an action for this condition from the options in the
"Corrective action(s)" box – see the table below.
3. Select the desired traffic light signal (YELLOW or GREEN, if
available) for the specified event by pressing the traffic light in the
"Traffic light signal" box – see the table below.
4. Select corrective actions/traffic light signal for the other conditions
in a similar way.
Conditions and corresponding corrective action options are as follows:
Condition Corrective action Traffic light
Calibration error(s)
present
Do not run scheduled built-in QC GREEN or
YELLOW
Calibration schedule
reminder(s)
Message on next patient result GREEN or
YELLOW
QC error(s) present "?" on specific parameters GREEN or
YELLOW
QC schedule
reminder(s)
Message on next patient result
Lock analyzer when QC overdue
GREEN or
YELLOW
GREEN or
YELLOW
Program
Conditions and
corrective
actions