ABL90 FLEX reference manual 1. Setup
1-59
QA Portal connection setup
This program allows you to connect the analyzer to a QA Portal.
If the QA Portal communication is enabled, the analyzer will automatically send
QC results and Cal Verification measurements to the QA Portal.
To enable communication with the QA Portal, do the following:
Step Action
1. Touch and highlight the "Server address" box.
Type in the TCP/IP address of your QA Portal, using the screen
keypad or keyboard.
2. Touch and highlight the "Port" box. Type in port number, using the
keypad.
3. Press the check button in the " QA Portal communication" box to
activate the connection.
The icon in the "Connection status" box indicates the state of the
QA Portal connection. "Connected" indicates an established
connection to the QA Portal.
4. Press Close to exit the screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals