ABL90 FLEX reference manual 7. Performance characteristics
7-7
Performance test results – chart description
General reference for the entire section: [Ref. 11].
Tests were performed in the following modes:
Mode Volume
Syringe 65 µL
Capillary 65 µL
The number of measurements during the test are listed below:
Parameter Number of
measurements
Test period,
days
Number of
Instruments
pH 786 30 8
pCO
2
691 25 10
pO
2
917 27 9
cK
+
800 13 10
cNa
+
799 30 8
cCa
2+
797 30 8
cCl
799 30 8
Blood: 1268 23 9 cGlu
Serum pool: 408
23 9
Blood: 1622 22 8 cLac
Serum pool: 546 22 8
ctHb 1042 8 10
sO
2
1193 22 10
FO
2
Hb 1193 22 10
FCOHb 1041 12 10
FMetHb 1049 9 10
FHHb 1193 22 10
FHbF 864 24 10
ctBil 649 16 10
Modes
Number of
measurements