ABL90 FLEX reference manual 7. Performance characteristics
7-7
Performance test results – chart description
General reference for the entire section: [Ref. 11].
Tests were performed in the following modes:
Mode Volume
Syringe 65 µL
Capillary 65 µL
The number of measurements during the test are listed below:
Parameter Number of
measurements
Test period,
days
Number of
Instruments
pH 786 30 8
pCO
2
691 25 10
pO
2
917 27 9
cK
+
800 13 10
cNa
+
799 30 8
cCa
2+
797 30 8
cCl
799 30 8
Blood: 1268 23 9 cGlu
Serum pool: 408
23 9
Blood: 1622 22 8 cLac
Serum pool: 546 22 8
ctHb 1042 8 10
sO
2
1193 22 10
FO
2
Hb 1193 22 10
FCOHb 1041 12 10
FMetHb 1049 9 10
FHHb 1193 22 10
FHbF 864 24 10
ctBil 649 16 10
Modes
Number of
measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals