ABL90 FLEX reference manual 7. Performance characteristics
7-17
Performance test results – FHHb
0%: whole blood is tonometered with a gas mixture containing 94.4% N
2
and 5.6% CO
2
+ dithionite.
Setup:
Adult samples. HbF correction is not activated.
FHHb
(%)
ctHb
(g/dL)
Bias
Sec.ref
S
0
S
x
CV
X
% TE
A
TE
A
(%)
0.0 15 0.07 0.10 0.28 - 0.62 -
10.0 15 0.08 0.18 0.35 3.5 0.77 7.7
25.0 15 0.05 0.30 0.48 1.9 0.99 4.0
35.0 15 0.08 0.27 0.50 1.4 1.06 3.0
50.0 15 0.11 0.26 0.57 1.1 1.23 2.5
100.0 15 -0.14 0.16 0.40 0.4 0.92 0.9
0.0 7 -0.45 0.13 0.36 - 1.16 -
0.0 25 0.53 0.09 0.26 - 1.04 -
Performance test results – FHbF
The reference method is based on Cation Exchange HPLC. The method is described in
[Ref. 15]. The method is performed by the Hæmatology Laboratory at Herlev Hospital,
Denmark.
Setup:
Mixed adult and Fetal samples. HbF correction enabled for all levels.
FHbF
(%)
ctHb
(g/dL)
Bias
Prim.ref
*)
S
0
S
X
CV
X
% TE
A
TE
A
(%)
0 15 1.6 1.8 3.0 - 7.48 -
5 15 3.2 1.8 3.3 66 9.67 193
10 15 0.9 1.7 3.2 32 7.17 71.7
20 15 1.1 1.8 3.6 18 8.16 40.8
30 15 -1.7 1.8 4.2 14 9.93 33.1
50 15 -3.1 1.6 3.9 7.8 10.74 21.5
80 15 -4.2 1.8 4.1 5.1 12.24 15.3
*) ABL735 corrected to HPLC through:
FHbF(ABL735),corr = FHbF (ABL735),meas – 0.9* ctHb + 11.7
Reference
method
Bias
Sec.ref
and
Repeatability –
blood samples
Reference
method
Bias
Prim.ref
and
Repeatability –
blood samples

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals