EasyManuals Logo

RADIOMETER ABL90 FLEX User Manual

RADIOMETER ABL90 FLEX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #203 background imageLoading...
Page #203 background image
ABL90 FLEX reference manual 7. Performance characteristics
7-17
Performance test results – FHHb
0%: whole blood is tonometered with a gas mixture containing 94.4% N
2
and 5.6% CO
2
+ dithionite.
Setup:
Adult samples. HbF correction is not activated.
FHHb
(%)
ctHb
(g/dL)
Bias
Sec.ref
S
0
S
x
CV
X
% TE
A
TE
A
(%)
0.0 15 0.07 0.10 0.28 - 0.62 -
10.0 15 0.08 0.18 0.35 3.5 0.77 7.7
25.0 15 0.05 0.30 0.48 1.9 0.99 4.0
35.0 15 0.08 0.27 0.50 1.4 1.06 3.0
50.0 15 0.11 0.26 0.57 1.1 1.23 2.5
100.0 15 -0.14 0.16 0.40 0.4 0.92 0.9
0.0 7 -0.45 0.13 0.36 - 1.16 -
0.0 25 0.53 0.09 0.26 - 1.04 -
Performance test results – FHbF
The reference method is based on Cation Exchange HPLC. The method is described in
[Ref. 15]. The method is performed by the Hæmatology Laboratory at Herlev Hospital,
Denmark.
Setup:
Mixed adult and Fetal samples. HbF correction enabled for all levels.
FHbF
(%)
ctHb
(g/dL)
Bias
Prim.ref
*)
S
0
S
X
CV
X
% TE
A
TE
A
(%)
0 15 1.6 1.8 3.0 - 7.48 -
5 15 3.2 1.8 3.3 66 9.67 193
10 15 0.9 1.7 3.2 32 7.17 71.7
20 15 1.1 1.8 3.6 18 8.16 40.8
30 15 -1.7 1.8 4.2 14 9.93 33.1
50 15 -3.1 1.6 3.9 7.8 10.74 21.5
80 15 -4.2 1.8 4.1 5.1 12.24 15.3
*) ABL735 corrected to HPLC through:
FHbF(ABL735),corr = FHbF (ABL735),meas – 0.9* ctHb + 11.7
Reference
method
Bias
Sec.ref
and
Repeatability –
blood samples
Reference
method
Bias
Prim.ref
and
Repeatability –
blood samples

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals