ABL90 FLEX reference manual 8. Parameters
8-39
The actual position of the ODC at 37 C for a given sample is, in principle,
determined in two steps:
1.
The calculation of the combined effect on the ODC position at 37 C of all
known causes for displacement (= ac in equation 46.13), and based on
this position.
2.
The computation by a numerical method of the actual position of the
ODC curve by shifting it to pass through the known set of coordinates
(P
0
, S
0
).
Eq. Description
46.13
a = ac + a6
46.14 ac = a1 + a2 + a3 + a4 + a5
46.15
a1 = 0.88 (pH 7.40)
46.16
2
CO
a2 0.048 ln
5.33
p

46.17
a3 0.7 MetHbF
46.18
a4 0.06 0.02 HbF DPG 5Fc
46.19
a5 0.25 HbFF
Step Description
I:
pO
2
, sO
2
can be used.
If sO
2
> 0.97, the calculation is based on II or III – see
below.
Coordinates (P
0
, S
0
) are calculated from equations
(46.9) and (46.11).
If FCOHb and FMetHb are not known, the default values
are used.
The ODC is shifted from the reference position to a
position that corresponds to the effect of all measured
parameters according to step I.
The magnitude of the shift is "ac".
The ODC is then further shifted to pass through the
point (P
0
, S
0
).
The magnitude of the shift is "a6".
The actual ODC
position
Determining
the actual
displacement

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing