5. Sensors and optical system ABL90 FLEX reference manual
5-46
The measured spectrum is compared to a model spectrum calculated from the
determined concentrations. The difference between these two spectra is called
the residual spectrum. If this residual spectrum is too high, the oximetry module
parameters
ctHb, sO
2
, FO
2
Hb, FCOHb, FMetHb, FHHb, FHbF and ctBil will be
flagged with a warning.
In addition, a warning will accompany the results if any of the following
conditions exist:
 ctHb < 0.1 mmol/L or ctHb > 25 mmol/L
 FHb(deriv) < –2 % or FHb(deriv) > 102 %
where
FHb(deriv) is defined as sO
2
, FO
2
Hb, FCOHb, FMetHb, FHHb
 SHb < 2 % or SHb > 10%
 Value of Turbidity < 0.5 % or > 5%
Residual
spectrum

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals