EasyManuals Logo

RADIOMETER ABL90 FLEX User Manual

RADIOMETER ABL90 FLEX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #174 background imageLoading...
Page #174 background image
5. Sensors and optical system ABL90 FLEX reference manual
5-46
The measured spectrum is compared to a model spectrum calculated from the
determined concentrations. The difference between these two spectra is called
the residual spectrum. If this residual spectrum is too high, the oximetry module
parameters
ctHb, sO
2
, FO
2
Hb, FCOHb, FMetHb, FHHb, FHbF and ctBil will be
flagged with a warning.
In addition, a warning will accompany the results if any of the following
conditions exist:
 ctHb < 0.1 mmol/L or ctHb > 25 mmol/L
 FHb(deriv) < –2 % or FHb(deriv) > 102 %
where
FHb(deriv) is defined as sO
2
, FO
2
Hb, FCOHb, FMetHb, FHHb
 SHb < 2 % or SHb > 10%
 Value of Turbidity < 0.5 % or > 5%
Residual
spectrum

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals