ABL90 FLEX reference manual 8. Parameters
8-13
Input parameters
Input parameters are the parameters keyed in by the operator on the Patient
Identification screen or transferred from an interfaced database.
All input parameters are given in this section.
Definition Patient temperature.
Unit °C; °F
Input range °C: 15.0-45.0
°F: 59-113
Conversion
T °F =
o
9
32
5
TC
; T C =
o
5
( 32)
9
TF
Definition Fraction of oxygen in dry inspired air.
Unit %; fraction
Input range %: 0-100
fraction: 0.000-1.000
Reference range 35.5-44.7 (m, f)
Is used if the analyzer version does not include the oximetry measuring system.
Definition Concentration of total hemoglobin in blood.
Input range /Unit g/dL: 0.0-33.0
g/L: 0-330
mmol/L: 0.0-20.5
Conversion
ctHb (g/dL) = 1.61140 ctHb (mmol/L);
ctHb (g/L) = 16.1140 ctHb (mmol/L);
ctHb (mmol/L) = 0.62058 ctHb (g/dL) =
0.062058 ctHb (g/L)
Definition Respiratory quotient, ratio between the CO
2
production and the O
2
consumption.
Input range 0.00-2.00
Definition Oxygen tension of mixed venous blood.
Input range/Unit mmHg; Torr: 0.0-750.0
kPa: 0.00-100
Conversion
p(kPa) = 0.133322 p(mmHg)
p(mmHg) = 7.500638 × p(kPa)
Definition
T
FO
2
(I)
ctHb
RQ
pO
2
(v
–
)