8. Parameters ABL90 FLEX reference manual
8-24
Eq. 5 [24]:
-
3
Base(B) (1 0.014 tHb)( HCO (P) 24.8 (1.43 tHb 7.7)(pH 7.4))ccc c
NOTICE: The formula is different from previous Radiometer analyzers. The
calculation is done in accordance with NCCLS (CLSI)-approved guidelines [24].
However, since it is assumed that the previous method [14] is a better model,
the previous range checks are retained (no values displayed outside ±50
mmol/L and values tagged with "?" outside the range of ±30 mmol/L).
The change corresponds to less than 0.6 mmol/L in the reference ranges for pH,
pCO
2
and ctHb.
Eq. 6 [4]:
)O1(tHb3062.0Base(B)ox)Base(B,
2
sccc
If ctHb is not measured or keyed in, the default value will be used.
If sO
2
is not measured, it will be calculated from equation 39.
Eq. 7 [24]:
-
3
Base(Ecf) HCO (P) 24.8 16.2 (pH 7.4)cc
See NOTICE in Eq. 5
Eq. 8:
cBase(Ecf,ox) = cBase(Ecf) – 0.3062 × 3 ×(1 – sO
2
)
Eq. 9 [4,14]:
-
3
HCO (P,st) 24.47 0.919 Z+Z a' Z-8c 
where
Eq. Description
9.1
-3 4
a'=4.04 10 4.25 10 tHbc

9.2
2
Base(B)-0.3062 tHb 1- OZc c s
Eq. 10 [4,5]:
-
223
tCO (P) 0.23 CO HCO (P)cpc
cBase(B)
cBase(B,ox)
cBase(Ecf)
cBase(Ecf,ox)
cHCO
3
(P,st)
ctCO
2
(P)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals