8. Parameters ABL90 FLEX reference manual
8-8
Can also be calculated: see Derived parameters eq. 39.
Definition Oxygen saturation, the ratio between the
concentrations of oxyhemoglobin and the hemoglobin
minus the dyshemoglobins.
Unit %; fraction
Range of indication %: –2-102
Fraction: –0.02-1.02
Reference range %: 95-99 (m, f)
Fraction: 0.95-0.99 (m, f)
Reference [11]
Can also be calculated: see Derived parameters eq. 40.
Definition Fraction of oxyhemoglobin in total hemoglobin in
blood.
Unit %; fraction
Range of indication %: –2-103
Fraction: –0.02-1.03
Reference range %: 94-98 (m, f)
Fraction: 0.94-0.98 (m, f)
Definition Fraction of carboxyhemoglobin in total hemoglobin in
blood.
Unit %; fraction
Range of indication %: –2-103
Fraction: –0.02-1.03
Reference range %: 0.5-1.5 (m, f)
Fraction: 0.005-0.015 (m, f)
Default value 0.004 (0.4 %)
Definition Fraction of methemoglobin in total hemoglobin in
blood.
Unit %; fraction
Range of indication %: –2-103
Fraction: –0.02-1.03
Reference range %: 0.0-1.5 (m, f)
Fraction: 0.000-0.015 (m, f)
Default value 0.004 (0.4 %)
sO
2
FO
2
Hb
FCOHb
FMetHb

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing