ABL90 FLEX reference manual 7. Performance characteristics
7-21
The following interference results are found for the electrolyte electrodes:
Interference on …....
Substance
Test con-
centration
cK
+
(4 mM
level)
cNa
+
(140 mM
level)
cCa
2+
(1.25 mM
level)
cCl
-
(105 mM
level)
Test
matrix
Lithium (Li
+
)
3.2 mM <0.1 <1 <0.02 N/A Plasma
Potassium (K
+
)
12 mM N/A <1 <0.02 N/A Plasma
3.4 mM
<0.1
1.2
N/A N/A Plasma
2.2 mM N/A
<1
N/A N/A Plasma
1.8 mM N/A
<1
N/A N/A Plasma
Calcium (Ca
2+
)
1.6 mM N/A
<1
N/A N/A Plasma
Sodium (Na
+
)
180 mM N/A N/A
0.029
N/A Plasma
1 mM
<0.1 <1
N/A
1.1
Plasma
Ammonium(NH
4
+
)
107 M <0.1 <1
N/A
<1
Plasma
Magnesium (Mg
2+
)
5 mM N/A <1 <0.02 N/A Aqueous
Zinc (Zn
2+
)
170 M <0.1 <1
0.024
N/A Plasma
Strontium (Sr
2+
)
150 M N/A N/A <0.02 N/A Plasma
N/A N/A
-0.037
mM/pH
N/A
Aqueous/
buffer
pH
6.8 – 8
N/A N/A N/A
<1
Plasma
37.5 mM N/A N/A N/A
76.6
Plasma
18.75 mM N/A N/A N/A
37.6
Plasma
10 mM N/A N/A N/A
19.5
Plasma
5 mM N/A N/A N/A
10.1
Plasma
Bromide (Br
-
)
1 mM N/A N/A N/A
1.8
Plasma
2.99 mM N/A N/A N/A
12.4
Plasma
1.5 mM N/A N/A N/A
5.3
Plasma
1 mM N/A N/A N/A
3.5
Plasma
Iodide (I
-
)
0.75 mM N/A N/A N/A
2.5
Plasma
107 M
N/A N/A N/A
<1
Plasma
Fluoride (F
-
)
1 mM N/A N/A N/A
<1
Plasma
0.375 mM N/A N/A
<0.02
2.1
Plasma
0.5 mM N/A N/A N/A
2.5
Plasma
0.75 mM N/A N/A N/A
3.7
Plasma
Perchlorate
(ClO
4
-
)
1.5 mM N/A N/A N/A
7.3
Plasma
7.5 g/mL
0.27 8.7 0.138
<1
Plasma
10 g/mL
0.39 12.1 0.182
<1
Plasma
15 g/mL
0.60 18.8 0.269
<1
Plasma
Benzalkonium chloride
30 g/mL
1.28 40.4 0.622
<1 Plasma
Electrolytes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals