ABL90 FLEX reference manual 8. Parameters
8-27
Eq. 18:
(I)O
),(aO
(I)O/)(a,O
2
2
22
F
Tp
FTp
The calculation cannot be performed on the basis of the default FO
2
(I) value,
and the calculation requires entering the sample as "Arterial" or "Capillary".
Eq. 19 Refer to Eq. 46.10:
The ODC is determined as described in equations 46-47 in the section
Oxyhemoglobin Dissociation Curve further in this chapter.

P
50=
COHb
1
0.5 1- COHb- MetHb
p
F
FF
where
Description See...
P = ODC(S,A,T) Eq. 47
0.5 1 COHb- MetHb COHb
1- MetHb
FF F
S
F

Eq. 46.11
A = a
T = 37.0 °C Eq. 46.13
Eq. 20:
The ODC is determined as described in equations 46-47 in the section
Oxyhemoglobin Dissociation Curve further in this chapter.

P
50( )=
COHb
1
0.5 1- COHb- MetHb
pT
F
FF
where
Description See…
P = ODC(S,A,T) Eq. 47
0.5 1 COHb- MetHb COHb
1- MetHb
FF F
S
F

Eq. 46.11

pH
a1.04 37.0
()
AT
T




23
pH
1.47 10 6.5 10 pH(37) 7.40
()T
T = patient temperature in °C (keyed-in)
pO
2
(a,T)/
FO
2
(I)
p50
p50(T)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing