EasyManuals Logo

RADIOMETER ABL90 FLEX User Manual

RADIOMETER ABL90 FLEX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #243 background imageLoading...
Page #243 background image
ABL90 FLEX reference manual 8. Parameters
8-27
Eq. 18:
(I)O
),(aO
(I)O/)(a,O
2
2
22
F
Tp
FTp
The calculation cannot be performed on the basis of the default FO
2
(I) value,
and the calculation requires entering the sample as "Arterial" or "Capillary".
Eq. 19 Refer to Eq. 46.10:
The ODC is determined as described in equations 46-47 in the section
Oxyhemoglobin Dissociation Curve further in this chapter.

P
50=
COHb
1
0.5 1- COHb- MetHb
p
F
FF
where
Description See...
P = ODC(S,A,T) Eq. 47
0.5 1 COHb- MetHb COHb
1- MetHb
FF F
S
F

Eq. 46.11
A = a
T = 37.0 °C Eq. 46.13
Eq. 20:
The ODC is determined as described in equations 46-47 in the section
Oxyhemoglobin Dissociation Curve further in this chapter.

P
50( )=
COHb
1
0.5 1- COHb- MetHb
pT
F
FF
where
Description See…
P = ODC(S,A,T) Eq. 47
0.5 1 COHb- MetHb COHb
1- MetHb
FF F
S
F

Eq. 46.11

pH
a1.04 37.0
()
AT
T




23
pH
1.47 10 6.5 10 pH(37) 7.40
()T
T = patient temperature in °C (keyed-in)
pO
2
(a,T)/
FO
2
(I)
p50
p50(T)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals