6. User-defined corrections ABL90 FLEX reference manual
6-4
Correction factors for pH and blood gases
The following corrections to the slope and offset are possible within the stated
limits for:
Arterial, venous, and a-v
samples:
Parameter Slope Offset
pH 0.80-1.20
0.05
pCO
2
0.80-1.20
10 mmHg
pO
2
0.80-1.20
20 mmHg
Correcting
slope and
offset