8. Parameters ABL90 FLEX reference manual
8-20
The following table shows the possible ranges and precision (number of
decimals) of the measured parameters. These ranges can be narrowed by
calculation ranges, reportable ranges, range of indication, etc., but should be
taken into consideration when external systems are interfaced to the analyzer.
Numerical format within the following ranges:
Symbol Unit
Range Range
pH(T) - 4.000 11.000
cH
+
(T) nmol/L -999999.0 199.9 200 9999999
pCO
2
(T) mmHg 0.0 99.9 100 750
kPa 0.00 9.99 10.0 100.0
cHCO
3
–
(P) mmol/L 0.0 100.0
cBase(B) mmol/L -50.0 50.0
cBase(B,ox) mmol/L -100.0 100.0
cBase(Ecf) mmol/L -50.0 50.0
cBase(Ecf,ox) mmol/L -100.0 100.0
cHCO
3
–
(P,st) mmol/L 0.0 150.0
ctCO
2
(P) mmol/L 0.0 100.0
Vol % 0.0 224.1
mL/dL 0.0 224.1
ctCO
2
(B) mmol/L 0.0 100.0
Vol % 0.0 224.1
mL/dL 0.0 224.1
pH(st) - 4.000 11.000
VCO
2
/V(dry
air)
% -10.0 110.0
fraction -0.100 1.100
Hct % -10.0 110.0
fraction -0.100 1.100
pO
2
(T) mmHg 0.0 99.9 100 750
kPa 0.00 9.99 10.0 100.0
pO
2
(A) mmHg 0.0 750.1
kPa 0.00 100.00
pO
2
(A,T) mmHg 0.0 750.1
kPa 0.00 100.00
p50 mmHg 0.00 750.06
kPa 0.00 100.00
p50(T) mmHg 0.00 750.06
kPa 0.00 100.00
p50(st) mmHg 0.00 750.06
kPa 0.00 100.00
Possible
ranges and
precision
(number of
decimals)