EasyManuals Logo

RADIOMETER ABL90 FLEX User Manual

RADIOMETER ABL90 FLEX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #237 background imageLoading...
Page #237 background image
ABL90 FLEX reference manual 8. Parameters
8-21
Numerical format within the following ranges
Symbol Unit
Range Range
pO
2
(Aa)
mmHg 0.0 750.1
kPa 0.00 100.00
pO
2
(Aa,T)
mmHg 0.0 750.1
kPa 0.00 100.00
pO
2
(a/A) % 0.0 10000.0
fraction 0.000 100.000
pO
2
(a/A, T) % 0.0 10000.0
fraction 0.000 100.000
pO
2
(a)/FO
2
(I) mmHg 0.0 99.9 100 7501
kPa 0.00 9.99 10.0 1000.0
pO
2
(a,T)/
FO
2
(I)
mmHg 0.0 99.9 100 7501
kPa 0.00 9.99 10.0 1000.0
pO
2
(x) mmHg 0.0 750.1
kPa 0.00 100.00
pO
2
(x,T) mmHg 0.0 750.1
kPa 0.00 100.00
ctO
2
(B) mmol/L 0.0 100.0
Vol % 0.0 224.1
mL/dL 0.0 224.1
ctO
2
(av
)
mmol/L 0.0 100.0
Vol % 0.0 224.1
mL/dL 0.0 224.1
BO
2
mmol/L 0.0 100.0
Vol % 0.0 224.1
mL/dL 0.0 224.1
ctO
2
(x) mmol/L 0.0 100.0
Vol % 0.0 224.1
mL/dL 0.0 224.1
D
·
O
2
mL/min 0 22414
mmol/min 0.0 1000.0
Q
·
t
L/min 0.0 100.0
V
·
O
2
mL/min 0 22414
mmol/min 0.0 1000.0

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals