EasyManuals Logo

RADIOMETER ABL90 FLEX User Manual

RADIOMETER ABL90 FLEX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #186 background imageLoading...
Page #186 background image
6. User-defined corrections ABL90 FLEX reference manual
6-8
Correction factors for electrolyte and metabolite
parameters
The following corrections to the slope and the offset are possible within the
stated limits for:
Blood samples:
Parameter Slope Offset
cK
+
0.80-1.20
1.0 mmol/L
cNa
+
0.80-1.20
10 mmol/L
cCa
2+
0.80-1.20
1.00 mmol/L
cCl
0.80-1.20
10 mmol/L
cGlu 0.750-1.250
5.0 mmol/L
cLac 0.750-1.250
5.0 mmol/L
The Radiometer default values for the electrolyte and metabolite parameters
must be reset manually by the user on the Parameter setup screen to 1.000
for each parameter.
Correcting
slope and
offset
Resetting
corrections to
default values

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals