6. User-defined corrections ABL90 FLEX reference manual
6-8
Correction factors for electrolyte and metabolite
parameters
The following corrections to the slope and the offset are possible within the
stated limits for:
Blood samples:
Parameter Slope Offset
cK
+
0.80-1.20
1.0 mmol/L
cNa
+
0.80-1.20
10 mmol/L
cCa
2+
0.80-1.20
1.00 mmol/L
cCl
0.80-1.20
10 mmol/L
cGlu 0.750-1.250
5.0 mmol/L
cLac 0.750-1.250
5.0 mmol/L
The Radiometer default values for the electrolyte and metabolite parameters
must be reset manually by the user on the Parameter setup screen to 1.000
for each parameter.
Correcting
slope and
offset
Resetting
corrections to
default values