1. Setup ABL90 FLEX reference manual
1-24
CAUTION – Changing QC
Changing a QC assigned to a slot will delete all current QC statistics
obtained on that slot. If you want a copy of the statistics for the last
QC month, create a WDC Report disk – see chapter 2: Disk functions
setup programs.
Quality control schedule setup
In this program you schedule QC measurements, both built-in and manual, for
your analyzer for all days of the week.
Navigation:
and
Use to select time during a day.
and
Use to display other weekdays.
Symbols for manual and built-in QC:
Measurement(s) on the manual QC.
Measurement(s) on the built-in QC performed automatically.