ABL90 FLEX reference manual 1. Setup
1-25
Adding a new QC solution to a schedule:
Step Action
1. Select the desired time and press Add to display the screen above.
2. Touch the "QC slot" box to activate it, if not already activated.
Select the desired slot/quality control solution, using the up/down
arrows in the box. Confirm with Select.
Built-in QC results are assigned to the slots A, B and C.
3. Highlight the "Week days" box using Field down and activate the
relevant check buttons to select the days of the week on which this
measurement should be performed.
4. Highlight the "Start time" box, using Field up, and key in the time
to perform a measurement and confirm with Enter on the keypad.
5. Highlight the "Repeat" box and select the interval with which the
measurement should be repeated, using the up/down arrows in the
box.
The QC schedule reminder "Lock analyzer when QC overdue"
(selected in Corrective actions – see further in this chapter) will
work on the basis of the setting selected in this box.
The symbols for the built-in or manual quality control will
automatically appear in the schedule.
6. Press OK to return to the Quality control schedule setup screen.