ABL90 FLEX reference manual 1. Setup
1-41
Parameters and input setup
Press Menu > Utilities > Setup > General setup > Parameters and input
setup and activate a button to enter a program.
The following programs are available:
 Parameters
 Units
 User-defined data items
 User-defined notes
Parameter setup
In this program you can do the following:
 Disable or enable a parameter
 Repress parameters if problems are detected
 Lock a parameter
 Make user-defined corrections for each measured parameter
 Make out-of-range suppression of oximetry parameters and bilirubin
Disabling/enabling a parameter:
Step Action
1. Highlight a parameter on the screen, using the scroll facilities.
2. Press Enable/Disable to include/exclude the parameter from a
parameter profile and the parameter bar. Note that pH, pCO
2
and
pO
2
cannot be excluded.
Program

Table of Contents

Related product manuals