ABL90 FLEX reference manual 5. Sensors and optical system
5-49
The uncorrected hemoglobin concentration,
ctHb(sample), measured on
capillary or syringe samples is corrected as follows:
cuv
tHb(sample)
tHb(sample,corr)
F
c
c
where:
ctHb(sample,corr)
= corrected ctHb
F
cuv
= analyzer-dependent cuvette path length constant
determined at tHb calibrations and automatically
stored by the analyzer
The uncorrected total bilirubin concentration,
ctBil(sample), measured on
capillary or syringe samples is corrected as follows:
cuv
tBil(sample)
tBil(sample,corr)
F
c
c
F
cuv
is the same as for tHb.
Corrections for
ctHb
Corrections for
ctBil