EasyManuals Logo

RADIOMETER ABL90 FLEX User Manual

RADIOMETER ABL90 FLEX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #205 background imageLoading...
Page #205 background image
ABL90 FLEX reference manual 7. Performance characteristics
7-19
For adult use: Adult samples within reference range:
The uncertainty in the bilirubin measurement on whole
blood can, in some cases, exceed the level required to
measure normal bilirubin levels for children older than 3
months and adults (bilirubin reference range
4-22 µmol/L).
Adult samples with an increased bilirubin level:
External tests using adult samples were performed on
samples with typically 80 % of the total bilirubin in the
conjugated form. For these highly conjugated samples
the external tests showed a negative bias of 18% on
whole-blood samples.
The patient samples represented typical variations in
ctBil, ctHb, sO
2
, pH and MCHC
(Mean Corpuscular Hemoglobin Concentration) values.
Three external tests were carried out at two different sites. Each test had its own ABL90
analyzer – a total of three.
Wet Chemistry analyzer Roche Modular with Roche Calibrator was used as a reference
[Ref. 21]
. Each external test site had two Modulars – a total of four. ctBil was measured
in
mol/L.
The field test results are given below:
S
yx
is the spreading around the linear line.
Actual external test from neonatal critical care hospitals using whole blood. Data from
three field tests are merged. Values are in µmol/L.
y = 1.014x - 0.828
R
2
= 0.985
0
50
100
150
200
250
300
350
400
0 100 200 300 400
M odu lar
ABL90
Syx=11.6
N=175

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals