EasyManuals Logo

RADIOMETER ABL90 FLEX User Manual

RADIOMETER ABL90 FLEX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #210 background imageLoading...
Page #210 background image
7. Performance characteristics ABL90 FLEX reference manual
7-24
Interference on
.......
Substance
Test
Concentration
Glu mM Lac mM
Test
matrix
2.5 mM
2.25
N/A
Whole blood
3.33 mM
2.88
N/A
Whole blood
5 mM
4.58
N/A
Whole blood
2-deoxy Glucose
10 mM
9.58
< 0.1
Whole blood
Unconjugated Bilirubin
0.2 g/L
< 0.1 < 0.1
Whole blood
Conjugated Bilirubin
0.2 g/L
< 0.1 < 0.1
Whole blood
Ascorbic acid
300 mg/dL
< 0.1 < 0.1
Whole blood