EasyManua.ls Logo

RADIOMETER ABL90 FLEX - Page 234

RADIOMETER ABL90 FLEX
346 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8. Parameters ABL90 FLEX reference manual
8-18
Symbol Definition Type Eq.
Q
·
t
Cardiac output; volume of blood delivered from
the left ventricle into the aorta per unit of time.
Also termed CO or C.O.
dv/in 32
V
·
O
2
Oxygen consumption; total amount of oxygen
utilized by the whole organism per unit of time.
dv/in 33
FO
2
(I) Fraction of oxygen in dry inspired air. in
FShunt Relative physiological shunt or concentration-
based shunt [5,8,9].
 Calculated from the pulmonary shunt
equation:
s
2
t
22
Q
1
tO (a v)
Q
1
tO (A) tO (a)
c
cc
if both arterial and mixed venous blood
samples are used.
 May be estimated from one arterial
sample by assuming a constant difference
in the concentrations of total oxygen in
arterial and mixed venous blood:
2
tO (a v) 2.3 mmol/L (5.15 mL/dL)c 
dv 34
FShunt (T) FShunt at patient temperature. dv 35
RI Respiratory Index; ratio between the oxygen
tension difference of alveolar air and arterial
blood and the oxygen tension of arterial blood.
dv 36
RI(T) Respiratory Index; ratio between the oxygen
tension difference of alveolar air and arterial
blood and the oxygen tension of arterial blood
at patient temperature.
dv 37
VO
2
/V(dry
air)
Volume fraction of oxygen in dry air. dv 52
Q
x
Cardiac oxygen compensation factor of arterial
blood defined as the factor by which the
cardiac output should increase to allow release
of 2.3 mmol/L (5.1 mL/dL) oxygen at a mixed
venous pO
2
of 5.0 kPa (38 mmHg) [5,8].
dv 38
V(B) Volume of blood, calculated when FCOHb and
V(CO) values are keyed in [5].
dv 42

Table of Contents