ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-9
No. Message Interpretation Operator action
378 Calibration
sensitivity out of
range
The sensitivity value is out of
range for the given
parameter.
- Check for and remedy
any system messages.
- Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration.
379 Calibration
unstable
(response fault)
An electrode response fault
occurred during calibration.
- Check for and remedy
any system messages.
- Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration.
443 Ca(7.4) not
usable
cCa
2+
at a pH of 7.4 is not
usable as the actual pH is
outside the 7.2-7.6 range.
- No action required.
452 Interference
during
measurement
Interference was detected
during measurement.
- Check the patient
record for medication
containing possible
interfering substances.
484 Today is last day
in stat. month -
remember to
print QC
statistics
After the current day, quality
control statistics obtained
over the month will be
deleted and new statistics
started.
- Print the QC statistics
if a copy is required.
487 A new statistical
month has begun
- remember to
export WDC data
A new statistical month has
begun.
- Make a WDC report
disk.
Removal condition:
- A WDC report disk has
been made.
494 Bilirubin too high
Detected bilirubin
concentration, ctBil(blood), is
greater than 2000 µmol/L.
The corresponding plasma
bilirubin concentration can be
calculated as follows:
ctBil(blood) = (1-Hct) ×
ctBil(plasma).
- No action required.