EasyManuals Logo

RADIOMETER ABL90 FLEX User Manual

RADIOMETER ABL90 FLEX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #317 background imageLoading...
Page #317 background image
ABL90 FLEX reference manual Appendix – Quality control
I-7
To correct the 2 SD control ranges for the above contributions, do the following:
Step Action
1.
Find the mean (X ) and the two times standard deviation (2 SD)
value in the Quality Control log (press: Data Logs > Quality
Control Log > Statistics) or calculate the values from the last 20
quality control measurements.
2. Find SD
total
in tables below.
3. Determine SD
corrected
as follows:

22
corrected total
SD 2 SD 2 SD
4.
Determine the user-defined control ranges as
corrected
XSD .
5. Enter the user-defined control ranges for each parameter in the
Control Ranges setup program (press:
Menu > Utilities > Setup
> QC Setup Control Ranges > Edit).
SD
total
for QUALICHECK5+ solutions:
Level 1
Parameter: ABL90 FLEX analyzer
pH 0.0041
pCO
2
kPa 0.14
pO
2
kPa
0.31
cK
+
mmol/L 0.035
cNa
+
mmol/L 0.7
cCa
2+
mmol/L 0.021
cCl
mmol/L
1.26
cGlu mmol/L 0.1
cLac mmol/L 0.1
ctBil mol/L
2.5
ctHb g/dL 0.11
FHbF % 1
Other Hb derivatives % 0.13
Level 2
Parameter: ABL90 FLEX analyzer
pH 0.0054
pCO
2
kPa 0.09
pO
2
kPa
0.26
cK
+
mmol/L 0.036
cNa
+
mmol/L 0.6
Table 1:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals