CHAPTER 10:
As a manual conditioning of the sensor cassette pack is not
possible for the time being, references to this option are no longer
valid and, therefore, the following analyzer messages no longer
exist in software version 2.4:
972 Sensor conditioning due
1080 Installed sensor cassette was not registered as conditioned
1333 Sensor conditioning started, no BC
Conditioning in
sensor cassette pack
not possible
1334 Sensor conditioning started, BC
Warm start With the new warm start feature in software version 2.4 the
following new analyzer message has been included:
No. Message Interpretation Operator action
1340 Sensor cassette
maintenance by
Analyzer has
been interrupted
This message is used in the Activity
log to indicate startup using a sensor
cassette which has been left without
an FTC activity for more than 1 hour.
No action required
TABLE OF CONTENTS/INDEX:
Conditioning in
sensor cassette pack
not possible
As the pack with the sensor cassette cannot be used for
conditioning, the name of the pack has been changed from
Conditioning unit to Sensor cassette pack.
The manual will be updated with the above information as part of the next manual
update.
Please place this note to the users in the binder of your manual.
Technical
documentation
Instructions to
user