1. Setup ABL90 FLEX reference manual
1-54
Step Action
1. Press Add.
2. Press Keyboard, type in the name of the connection instead of the
default one, and press Enter.
Press Back to return to the LIS/HIS connection setup screen.
3. Select the high-level protocol according to the requirements of the
connected device, using the up/down arrows in the box.
Available protocols:
ASTM, ASTM6xx, HL7 ver. 2.2, HL7 version 2.5 or POCTDML1A.
4. Select the low-level protocol as follows:
 Use "Serial" or "Serial (RAW)" for the serial connection
 Use "Network (TCP/IP)" for the network connection
 Use "Network (TCP/IP)ASTM" for additional serial connection (not
all combinations of high-level protocols and low-level protocols
are possible)