Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1054User Manual 1178.8520.02 ─ 01
To query the limit check result, use CALCulate<n>:LIMit<li>:FAIL?.
Note that a new command exists to activate the limit check and define the trace to be
checked in one step (see CALCulate<n>:LIMit<li>:TRACe<t>:CHECk
on page 1054).
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:LIM:STAT ON
Switches on the limit check for limit line 1.
Manual operation: See " Disable All Lines " on page 477
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk <State>
This command turns the limit check for a specific trace on and off.
To query the limit check result, use CALCulate<n>:LIMit<li>:FAIL?.
Note that this command replaces the two commands from previous signal and spec-
trum analyzers (which are still supported, however):
CALCulate<n>:LIMit<li>:TRACe<t> on page 1196
CALCulate<n>:LIMit<li>:STATe on page 1053
Suffix:
<n>
.
Window
<li> Limit line
<t> Trace
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:LIM3:TRAC2:CHEC ON
Switches on the limit check for limit line 3 on trace 2.
Manual operation: See " Traces to be Checked " on page 476
Analyzing Measurements (Basics)